Hospira brand Symbiq Two-Channel Infuser, LIST No. 16027 Symbiq Two-Channel Infuser, List Number Configurations: 16027-04-01/03/04, 51/52, 53/54, 79/80,81/82,83/84,87/88 and 89/90 16026-27-89/90, 16027-27-89/90, 16027-13-25/26,53/54, Distributed by: Hospira, Inc. Morgan Hill, CA

This recall has been terminated (originally issued September 9, 2010).

Company
Hospira Inc
Recall Initiated
February 22, 2010
Posted
September 9, 2010
Terminated
April 1, 2011
Recall Number
Z-2391-2010
Firm Location
Morgan Hill, CA
Official Source
View on FDA website ↗

Reason for Recall

The clamp may not secure to the IV pole and the pump may slip or fall off. The pump may fall on a person or pull the tubing out of the IV container.

Distribution

Worldwide Distribution: USA (all states and Puerto Rico) and the countries of Australia, Canada, Colombia, Costa Rica, Malaysia, Mexico, Singapore. and the UAE.

Lot / Code Info

All units of these List Numbers.

Root Cause

Device Design

Action Taken

Hospira issued an "Urgent: Device Field Correction" Letter dated February 22, 2010 to all consignees, identifying the reason for the recall and affected product. Customers were informed that they will be contacted by a Hospira representative to replace the affected pole clamps. Consignees can contact Hospira Technical Support Operations at 1-800-241-4002.

More recalls by Hospira Inc