Hospira Inc. Symbiq One- and Two- Channel Infuser, List Numbers 16026 and 16027 respectively.

This recall has been terminated (originally issued September 8, 2010).

Company
Hospira Inc
Recall Initiated
March 26, 2010
Posted
September 8, 2010
Terminated
March 30, 2012
Recall Number
Z-2381-2010
Quantity
28485 units
Firm Location
Morgan Hill, CA
Official Source
View on FDA website ↗

Reason for Recall

If the administrator set is removed prior to the cassette carriage fully opening without closing the slide/roller clamp, unrestricted flow and delivery may result.

Distribution

Worldwide distribution: USA, Australia, Canada, Colombia, Costa Rica, Malaysia, Mexico, Puerto Rico, Singapore and the UAE.

Lot / Code Info

List number configurations: 16026-04-01/03/04, 51/52, 53/54, 79/80, 81/82, 83/84, 87/88, 89/90; 16027-04-01/03/04, 51/52, 53/54, 79/80, 81/82, 83/84, 87/88, 89/90; 16026-27-89/90, 16026-13-25/26, 53/54; 16027-27-89/90; 16027-13-25/26, 53/54.

Root Cause

Device Design

Action Taken

Hospitra Urgent Device Correction Notification mailed on 03/26/2010 via federal express. Tracking done by delivery tracking. Distribution of product continues and has not stopped. Additional notification was issued in 09/2010. A software correction is planned.