Hospira brand Symbiq One-Channel Infuser LIST No. 16026 Symbiq One-Channel Infuser, List Number Configurations: 16026-04-01/03/04,51/52, 53/54, 79/80, 81/82, 83/84, 87/88 and 89/90, 16026-13-25/26, 53/54, Distributed by: Hospira, Inc. Morgan Hill, CA
This recall has been terminated (originally issued September 9, 2010).
- Company
- Hospira Inc
- Recall Initiated
- February 22, 2010
- Posted
- September 9, 2010
- Terminated
- April 1, 2011
- Recall Number
- Z-2390-2010
- Firm Location
- Morgan Hill, CA
- Official Source
- View on FDA website ↗
Reason for Recall
The clamp may not secure to the IV pole and the pump may slip or fall off. The pump may fall on a person or pull the tubing out of the IV container.
Distribution
Worldwide Distribution: USA (all states and Puerto Rico) and the countries of Australia, Canada, Colombia, Costa Rica, Malaysia, Mexico, Singapore. and the UAE.
Lot / Code Info
All units of these List Numbers.
Root Cause
Device Design
Action Taken
Hospira issued an "Urgent: Device Field Correction" Letter dated February 22, 2010 to all consignees, identifying the reason for the recall and affected product. Customers were informed that they will be contacted by a Hospira representative to replace the affected pole clamps. Consignees can contact Hospira Technical Support Operations at 1-800-241-4002.