Hospira Power cord use on Symbiq One-Channel infuser and Symbiq Two-Channel infuser. Power cord part number: HSP3306-E07. Used on List No. 16026 - Symbiq One Channel infuser; List No. 16027 - Symbiq Two-Channel infuser
This recall has been terminated (originally issued March 8, 2011).
- Company
- Hospira Inc
- Recall Initiated
- October 27, 2010
- Posted
- March 8, 2011
- Terminated
- November 7, 2013
- Recall Number
- Z-1596-2011
- Quantity
- 32,376 power cords
- Firm Location
- Morgan Hill, CA
- Official Source
- View on FDA website ↗
Reason for Recall
Complaints of broken, bent or missing prongs, charring, sparks, visible smoke, burnt smell have been reported on Symbiq AC Power cords.
Distribution
Nationwide distribution and to Australia, Canada, and Malaysia.
Lot / Code Info
Power cord part number: HSP3306-E07. Used on List No. 16026 - Symbiq One Channel infuser; List No. 16027 - Symbiq Two-Channel infuser List number configurations: 16026-04-01/03/04, 51/52,. 53/54, 79/80, 81/82, 83/84/ 87/88; 16027-04-01/03/04, 51/52, 53/54, 79/80, 81/82, 83/84, 87/88; 16026-13-25/26, 53/54; 16027-13-25/26, 53/54.
Root Cause
Component design/selection
Action Taken
Hospira Recall notification letters were sent by Federal Express on November 1, 2010.