Hospira Power cord use on Symbiq One-Channel infuser and Symbiq Two-Channel infuser. Power cord part number: HSP3306-E07. Used on List No. 16026 - Symbiq One Channel infuser; List No. 16027 - Symbiq Two-Channel infuser

This recall has been terminated (originally issued March 8, 2011).

Company
Hospira Inc
Recall Initiated
October 27, 2010
Posted
March 8, 2011
Terminated
November 7, 2013
Recall Number
Z-1596-2011
Quantity
32,376 power cords
Firm Location
Morgan Hill, CA
Official Source
View on FDA website ↗

Reason for Recall

Complaints of broken, bent or missing prongs, charring, sparks, visible smoke, burnt smell have been reported on Symbiq AC Power cords.

Distribution

Nationwide distribution and to Australia, Canada, and Malaysia.

Lot / Code Info

Power cord part number: HSP3306-E07. Used on List No. 16026 - Symbiq One Channel infuser; List No. 16027 - Symbiq Two-Channel infuser List number configurations: 16026-04-01/03/04, 51/52,. 53/54, 79/80, 81/82, 83/84/ 87/88; 16027-04-01/03/04, 51/52, 53/54, 79/80, 81/82, 83/84, 87/88; 16026-13-25/26, 53/54; 16027-13-25/26, 53/54.

Root Cause

Component design/selection

Action Taken

Hospira Recall notification letters were sent by Federal Express on November 1, 2010.