Hospira brand Symbiq One-Channel Infuser LIST No. 16026 Symbiq One-Channel Infuser, List Number Configurations: 16026-04-01/03/04,51/52, 53/54, 79/80, 81/82, 83/84, 87/88 and 89/90, 16026-13-25/26, 53/54, Distributed by: Hospira, Inc. Morgan Hill, CA

This recall has been terminated (originally issued October 22, 2010).

Company
Hospira Inc
Recall Initiated
February 22, 2010
Posted
October 22, 2010
Terminated
December 3, 2013
Recall Number
Z-0127-2011
Firm Location
Morgan Hill, CA
Official Source
View on FDA website ↗

Reason for Recall

Hospira has identified motor encoder failures in the Symbiq pumping mechanism that causes the infuser to cease operation during infusion therapy, resulting in delay or interruption of infusion therapy.

Distribution

Worldwide Distribution: USA and the coutnries of Australia, Canada, Colombia, Costa Rica, Malaysia, Mexico, Puerto Rico, Singapore and the UAE.

Lot / Code Info

All units of these List Numbers.

Root Cause

Device Design

Action Taken

Hospira issued a Recall Notification letter dated October 4, 2010, updating the Urgent Device Field Correction letter dated February 22, 2010, ideentifying the affected products, the issue prompting the recall, and actions to be taken by customers. Customers are to examine their inventory to determine if they have any of the affected products, and recommended to remove them from service in critical care areas, and contact Hospira to schedule removal and replacement of the motor. Replace affected devices with a loaner or corrected back-up pump if available. Custoemrs can contacvtd Hospira Global Product Safety and Complaints at 1-800-441-4100, and Technical Support Operations at 1-800-241-4002.