Hospira Phoenix Infusion System with MedNet Software, Symbiq Two-Channel Infusion System, Software versions v2.1 & v3.0, Manufactured by Hospira Inc., Morgan Hill, CA

This recall has been terminated (originally issued August 11, 2009).

Company
Hospira Inc
Recall Initiated
March 31, 2009
Posted
August 11, 2009
Terminated
December 18, 2010
Recall Number
Z-1816-2009
Firm Location
Morgan Hill, CA
Official Source
View on FDA website ↗

Reason for Recall

Potential delay/underinfusion of critical therapy-- Devices experience increased frequency of alarms that result in temporary pump stoppage.

Distribution

Product was distributed to 59 consignees to all 50 States.

Lot / Code Info

List number 16027-04-83/84 and 87/88, software versions 2.1 and 3.0 only.

Root Cause

Software design

Action Taken

Hospira, Inc. issued an "Urgent: Device Field Correction" dated March 31, 2009 via Federal Express to Healthcare Professionals and Valued Hospira Customers. Consignees were informed of the affected devices and will be contacted by a Hospira representative to arrange necessary device upgrades. For further questions, contact Hospira Global Product Safety and Complaints at 1-800-441-4100.