Plum A+3 Infusion System with Hospira Mednet, list # 20678-04-01, 20678-04-03, 20678-04-51, 20678-04-77, 20678-04-81, 20678-04-05, 20678-04-53; AC cord manufactured by Electri-Cord, Westfield PA, Infusion pumps manufactured by Hospira, Morgan Hill, CA.

This recall has been terminated (originally issued October 6, 2009).

Company
Hospira Inc
Recall Initiated
August 11, 2009
Posted
October 6, 2009
Terminated
November 30, 2010
Recall Number
Z-0027-2010
Quantity
204,999 total
Firm Location
Morgan Hill, CA
Official Source
View on FDA website ↗

Reason for Recall

Fire/Shock hazard-- A component used in the device may cause sparks/flashes and poses a fire and shock hazard.

Distribution

Worldwide Distribution

Lot / Code Info

AC power cords Part Numbers: 826-85559-001,  826-85559-003,  826-85559-008,  826-85559-010, and  826-96057-003.

Root Cause

Component design/selection

Action Taken

Firm initiated its recall on 08/14/2009. A nationwide press release was issued to AP, and an Urgent: Device Field Correction letter to consignees was sent via UPS. The letter recommended that customers check their infusion pumps and spare parts in their inventory to determine if the specified AC Power Cords are in use at their facility; and, to inspect the affected cords for bent or cracked prongs and bridge connection, burnt plastic, or excessive wear and tear. Also, it was recommended that customers immediately discontinue use of the affected cords exhibiting the described characteristics and contact their local Hospira representative for assistance. The letter stated that customers may continue to use and monitor cords that are not exhibiting the described characteristics until replacement cords are sent to their facility. Hospira will contact customers to schedule replacement. Customers are to complete and return the attached Reply Form. Question should be directed to Hospira Customer Service at 1-800-241-4002.