Hemoliance RecombiPlasTin. Prothrombin Time Test. The product is available as a kit of 25 vials of Hemoliance RecombiPlasTin matched with 25 vials of Hemoliance RecombiPlasTin Diluent. Hemoliance RecombiPlasTin after reconstitution with Hemoliance RecombiPlasTin Diluent is a liposomal preparation of recombinant human tissue factor and purified phospholipids, calcium chloride, buffer and a preservative. The use of recombinant human tissue factor and purified phospholipids assures uniformity in reagent purity, ISI values and performance from lot to lot. Once reconstituted, it is ready for use in the one-stage PT test. Hemoliance RecombiPlasTin Diluent is an aqueous solution of calcium chloride and a preservative. --- FOR IN VITRO DIAGNOSTIC USE.

This recall has been terminated (originally issued July 18, 2006).

Company
Instrumentation Laboratory Co.
Recall Initiated
March 7, 2006
Posted
July 18, 2006
Terminated
August 11, 2006
Recall Number
Z-1247-06
Quantity
5 mL: 3870; 20 mL: 29,007 (all Domestic)
Firm Location
Lexington, MA
Official Source
View on FDA website ↗

Reason for Recall

Firm received 2 complaints concerning incorrect prothrombin time (PT) results while using the reagents on samples of patients taking the antibiotic CUBICIN (Daptomycin for injection).

Distribution

Nationwide

Lot / Code Info

Product No.49732750 (5 mL): Lots N0448239, N0549190, N0940581, N0353406, N0857200, N1050152; Product No. 49732720 (20 mL): Lots N0347895, N0548959, N0749495, N0940543, N0940586, N1242409, N0353410, N0454257, N0857199, N1050143.

Root Cause

Other

Action Taken

"Urgent Product Notifications" were sent to IL customers on 3/07/06 through its distributor, Beckman Coulter.