GEM Premier 4000 Laboratory Analyzer

This recall has been terminated (originally issued May 3, 2007).

Company
Instrumentation Laboratory Co.
Recall Initiated
March 27, 2007
Posted
May 3, 2007
Terminated
May 23, 2012
Recall Number
Z-0778-2007
Quantity
210
Firm Location
Lexington, MA
Official Source
View on FDA website ↗

Reason for Recall

Instrument in Micro Sampling Mode failed to aspirate and no error code was generated and an incorrect patient report could result

Distribution

Worldwide, including USA, Australia, Austria, Greece, Spain, Netherlands, Great Britain, France, Germany, Italy, and Sweden

Lot / Code Info

Serial numbers prior to 07020343 with Software versions prior to V1.1.3

Root Cause

Other

Action Taken

Instrumentation Labs on 3/2707 notified Service Reps to perform on site mandatory upgrade. The foreign counterparts were advised by email on 3/21/07