Instrumentation Laboratory ACL TOP Coagulation Analyzer (Base Model only). Blood analyzer. P/N 280000
This recall has been terminated (originally issued August 7, 2007).
- Company
- Instrumentation Laboratory Co.
- Recall Initiated
- June 19, 2007
- Posted
- August 7, 2007
- Terminated
- May 23, 2012
- Recall Number
- Z-1127-2007
- Quantity
- 94
- Firm Location
- Lexington, MA
- Official Source
- View on FDA website ↗
Reason for Recall
ACL TOP unit, when in emergency stop may release liquid during recovery sequence and contaminate sample tube.
Distribution
Nationwide.
Lot / Code Info
Serial numbers prior to 07060782
Root Cause
Other
Action Taken
Instrumentation Laboratory notiifed the distributor Beckman Coulter on 6/19/2007. Beckman Coulter notified their customers and provided a copy with the Instructions for Recovery Procedure on how to avoid contamination from probe leaks.