Instrumentation Laboratory ACL TOP Coagulation Analyzer (Base Model only). Blood analyzer. P/N 280000

This recall has been terminated (originally issued August 7, 2007).

Company
Instrumentation Laboratory Co.
Recall Initiated
June 19, 2007
Posted
August 7, 2007
Terminated
May 23, 2012
Recall Number
Z-1127-2007
Quantity
94
Firm Location
Lexington, MA
Official Source
View on FDA website ↗

Reason for Recall

ACL TOP unit, when in emergency stop may release liquid during recovery sequence and contaminate sample tube.

Distribution

Nationwide.

Lot / Code Info

Serial numbers prior to 07060782

Root Cause

Other

Action Taken

Instrumentation Laboratory notiifed the distributor Beckman Coulter on 6/19/2007. Beckman Coulter notified their customers and provided a copy with the Instructions for Recovery Procedure on how to avoid contamination from probe leaks.