ACL Futura, ACL Advance and ACL TOP Coagulation Analyzers used with the HemosIL RecombiPlasTin reagent

This recall has been terminated (originally issued June 21, 2005).

Company
Instrumentation Laboratory Co.
Recall Initiated
May 19, 2005
Posted
June 21, 2005
Terminated
December 6, 2006
Recall Number
Z-0915-05
Quantity
617 units
Firm Location
Lexington, MA
Official Source
View on FDA website ↗

Reason for Recall

Patient prothrombin time (PT) may report low for individuals on anticoaglant therapy

Distribution

Nationwide Canada

Lot / Code Info

Models:  290001 999001 with SN''s beginning with 05041164 280000 with S/N''s beginning with 0505354

Root Cause

Other

Action Taken

Instrumentation Laboratory issued an 'Urgent Product Notification' on 5/19/2005 to Beckman Coulter, Miami FL (DIstributor) to provide the written notification and advise customers to review PT curves an any patient on anticagulant treatment who reports a low result , and manually view the curve where a PT is flagged. A software upgrade will be availabel in July 2005.