Navitrack System - OS Knee Universal, Orthopedic Stereotaxic Instrument CAS Software application intended to assist in the placement of total knee replacement components

This recall has been terminated (originally issued April 26, 2017).

Company
Orthosoft, Inc. Dba Zimmer Cas
Recall Initiated
October 6, 2011
Posted
April 26, 2017
Terminated
May 8, 2017
Recall Number
Z-1881-2017
Quantity
105
Official Source
View on FDA website ↗

Reason for Recall

Zimmer CAS voluntarily conducted a retrospective recall of the Navitrack System - OS Knee Universal software ORTHOsoft-UniTkr-2.3.2.6, due to a calibration sequence crash.

Distribution

Worldwide Distribution - US including AZ CA CO FL GA IL IN KS MD MI MN MS OH OR PA SC SD TX VA WA and WI Internationally to Canada, Australia Austria Colombia Denmark France Germany Korea New Zealand Russia South Africa and Thailand

Lot / Code Info

Part Name: OS Knee Universal Part Number: ORTHOsoft-UniTkr-2 Affected Lot: ORTHOsoft-UniTkr-2.3.2.6

Root Cause

Software design

Action Taken

Zimmer sent a Medical Device Field Notification letter dated December 6, 2011, to all affected customers. This is a retrospective report of a correction taken on 6 December 2011. The software issues were fixed in the new software version 2.3.3.0. Recall notices were distributed on December 6, 2011. Telephone: 18663367846 (US & Canada) or 5143958883 (International), Email: cassupport@zimmercas.com Customers with questions should call 1-866-336-7846 (US & Canada) 514-395-8883 (International). For questions regarding this recall call 574-372-4487.