NDI P7 Position Sensor, Stereotaxic Instrument
This recall has been terminated (originally issued May 8, 2017).
- Company
- Orthosoft, Inc. Dba Zimmer Cas
- Recall Initiated
- June 9, 2009
- Posted
- May 8, 2017
- Terminated
- May 9, 2017
- Recall Number
- Z-2048-2017
- Quantity
- 34
- Official Source
- View on FDA website ↗
Reason for Recall
Zimmer CAS voluntarily conducted a retrospective recall on specific P7 Position Sensors (Cameras), installed on Sesamoid and Sesamoid Plasty CAS workstations due to a series of these components determined to be affected with a manufacturing issue whereby they may potentially stop functioning during usage.
Distribution
Worldwide Distribution - US Distribution to the states of : KS, MS, WI, NJ, KS, IN, LA, FL and WA., and to the countries of : Canada, Australia, Singapore, Netherlands, UK, Austria, Germany, Switzerland and Czech Republic.
Lot / Code Info
Part Number: NDI P7 Position Sensor Product Number 201.053 (a component of part number 521.035, Sesamoid Plasty Camera Arm Kit) Lot/Serial Number: 38 units affected: P7-00321, P7-00339, P7-00360, P7-00361, P7-00362, P7-00363, P7-00364, P7-00365, P7-00366, P7-00368, P7-00369, P7-00370, P7-00371, P7-00372, P7-00373, P7-00375, P7-00387, P7-00412, P7-00414, P7-00415, P7-00416, P7-00417, P7-00418, P7-00419, P7-00420, P7-00421, P7-00422, P7-00423, P7-00431, P7-00451, P7-00452, P7-00453, P7-00562, P7-00563, P7-00565, P7-00568, P7-00569, P7-00590 (The lot number and the serial number are the same)
Root Cause
Nonconforming Material/Component
Action Taken
Zimmer CAS sent an Field Recall Notice dated June 9, 2009. Zimmer CAS conducted a retrospective recall/Field Correction on given P7 Position Sensors (Cameras). These were installed on given Sesamoid and Sesamoid Plasty CAS workstations. A series of these components have been determined to be affected with a manufacturing issue whereby they may potentially stop functioning during usage. The failure mode consists in the camera stopping to track and interrupting the application with an error code 0X00000020 being displayed on the navigation panel. For any questions or concerns please contact the Zimmer CAS customer service, 1-866-336-7846.