NDI P7 Position Sensor, Stereotaxic Instrument

This recall has been terminated (originally issued May 8, 2017).

Company
Orthosoft, Inc. Dba Zimmer Cas
Recall Initiated
June 9, 2009
Posted
May 8, 2017
Terminated
May 9, 2017
Recall Number
Z-2048-2017
Quantity
34
Official Source
View on FDA website ↗

Reason for Recall

Zimmer CAS voluntarily conducted a retrospective recall on specific P7 Position Sensors (Cameras), installed on Sesamoid and Sesamoid Plasty CAS workstations due to a series of these components determined to be affected with a manufacturing issue whereby they may potentially stop functioning during usage.

Distribution

Worldwide Distribution - US Distribution to the states of : KS, MS, WI, NJ, KS, IN, LA, FL and WA., and to the countries of : Canada, Australia, Singapore, Netherlands, UK, Austria, Germany, Switzerland and Czech Republic.

Lot / Code Info

Part Number: NDI P7 Position Sensor Product Number 201.053 (a component of part number 521.035, Sesamoid Plasty Camera Arm Kit) Lot/Serial Number: 38 units affected: P7-00321, P7-00339, P7-00360, P7-00361, P7-00362, P7-00363, P7-00364, P7-00365, P7-00366, P7-00368, P7-00369, P7-00370, P7-00371, P7-00372, P7-00373, P7-00375, P7-00387, P7-00412, P7-00414, P7-00415, P7-00416, P7-00417, P7-00418, P7-00419, P7-00420, P7-00421, P7-00422, P7-00423, P7-00431, P7-00451, P7-00452, P7-00453, P7-00562, P7-00563, P7-00565, P7-00568, P7-00569, P7-00590 (The lot number and the serial number are the same)

Root Cause

Nonconforming Material/Component

Action Taken

Zimmer CAS sent an Field Recall Notice dated June 9, 2009. Zimmer CAS conducted a retrospective recall/Field Correction on given P7 Position Sensors (Cameras). These were installed on given Sesamoid and Sesamoid Plasty CAS workstations. A series of these components have been determined to be affected with a manufacturing issue whereby they may potentially stop functioning during usage. The failure mode consists in the camera stopping to track and interrupting the application with an error code 0X00000020 being displayed on the navigation panel. For any questions or concerns please contact the Zimmer CAS customer service, 1-866-336-7846.