Navitrack System - OS Knee Universal, Orthopedic Stereotaxic Instrument

This recall has been terminated (originally issued May 12, 2017).

Company
Orthosoft, Inc. Dba Zimmer Cas
Recall Initiated
June 20, 2008
Posted
May 12, 2017
Terminated
May 15, 2017
Recall Number
Z-2075-2017
Quantity
152
Official Source
View on FDA website ↗

Reason for Recall

ZIMMER CAS voluntarily conducted a retrospective recall of the Navitrack System - OS Knee Universal, due to software anomalies which were corrected with a new revision in version 2.2.1.3.

Distribution

Domestic: CA, FL, GA, IL, IN, KS, MD, MN, MS, NC, NJ, NY, OR, PA, RI, SD, TN, TX, WA, WI Foreign: Germany, Spain, Czech Republic, Netherlands, Switzerland, New Zealand, Morocco, Brazil, Australia, Austria, France, Italy, Singapore, United Kingdom

Lot / Code Info

Product Name: Orthosoft Universal TKR (software application) Product Number: ORTHOsoft-UniTkr-2 Affected Lot #: Software release version ORTHOsoft-UniTkr-2.2.0.82

Root Cause

Software design

Action Taken

This is a retrospective report of a correction taken on 20 June 2008. The two software issues were corrected in the new software version 2.2.1.3. There were 164 copies of the affected software, 5 of which were only temporary licenses. There were 152 of the 159 software copies that had a full license that were confirmed to have installed the updated instructions per the correction notice for a total percentage of 96%. All sales representatives were made aware of the software issues by written notice. Both software issues were corrected in a new revision of the software with no remaining effect. There have been no complaints since the correction was implemented. Recall notices were e-mailed to customers dated 06/20/2008.