Navitrack System - OS Knee Universal, Stereotaxic instrument

This recall has been terminated (originally issued May 12, 2017).

Company
Orthosoft, Inc. Dba Zimmer Cas
Recall Initiated
October 28, 2008
Posted
May 12, 2017
Terminated
May 15, 2017
Recall Number
Z-2076-2017
Quantity
153
Official Source
View on FDA website ↗

Reason for Recall

Zimmer CAS voluntarily conducted a retrospective recall of the Navitrack System OS Knee Universal software versions 2.2.0.82, 2.2.1.3, 2.1.6.0.2, D.54 and A.42 due to software issues resulting in erros and discrepancies .

Distribution

Domestic: CA, FL, GA, IL, IN, KS, NC, NY, OH, OR, PA, RI, TN, TX, VA, WA Foreign: Canada, Australia Austria Brazil China Czech Republic France Germany Israel Italy Lebanon Luxembourg Netherlands New Zealand Spain Sweden Switzerland United Kingdom

Lot / Code Info

Part Number, Lot Number/Serial Number ORTHOsoft-UniTkr- 2.2.0.82, 2.2.1.3 ORTHOsoft-UniTkr- 2.1.6.0.2 OsUniTkr (previous naming convention for ORTHOsoft-UniTkr)- D.54, A.42

Root Cause

Software design

Action Taken

This is a retrospective report of a correction taken on 28 October, 2008. The software issues mentioned above were corrected in subsequent software updates. There were 153 units of the affected software distributed. There were 147 of the 153 units confirmed to have installed the updated software per the correction notice for a total percentage of 96%. Recall Notices were sent to customers on 10/28/2008.