Omega Medical Imaging Elevating Monitor Suspension Systems, Part Number 1000-0085

This recall is currently active, issued October 24, 2023. It was issued by Omega Medical Imaging, Inc..

Company
Omega Medical Imaging, Inc.
Recall Initiated
September 8, 2023
Posted
October 24, 2023
Recall Number
Z-0160-2024
Quantity
31 units
Firm Location
Sanford, FL
Official Source
View on FDA website ↗

Reason for Recall

An actuator separated from the pivot mechanism on a lower monitor boom system column. The lower boom became detached at the point shown and was restrained by the installed safety cable.

Distribution

Alabama, California, Illinois, Indiana, Kentucky, Louisiana, Missouri, Montana, North Carolina, Ohio, Oklahoma

Lot / Code Info

all units manufactured between 10-22-2012 and 07-09-2019

Root Cause

Device Design

Action Taken

Omega Medical Imaging issued a Field Safety Advisory Notice to its consignees on 09/08/2023 via email. The notice explained the problem and the risk. The firm will inspect all affected devices to ensure the proper fasteners were installed on the unit.