RecallDepth

Omega Medical Imaging - 2 over 2 Elevating Monitor Suspension pillar/column accessory Device accessory that elevates monitor for healthcare professionals.

This recall has been terminated (originally issued June 19, 2017).

Company
Omega Medical Imaging, Inc.
Recall Initiated
March 23, 2017
Posted
June 19, 2017
Terminated
December 11, 2019
Recall Number
Z-2621-2017
Quantity
6 units
Firm Location
Sanford, FL
Official Source
View on FDA website ↗

Reason for Recall

Elevating Monitor Suspension column dropping and will not rise back up.

Distribution

Worldwide Distribution -US Distribution to states of: IN, OH, PA, and MT; and country of: Pakistan.

Lot / Code Info

SKF Telescopic Pillar P/N ending in -003

Root Cause

Mixed-up of materials/components

Action Taken

The firm, OMEGA MEDICAL IMAGING, sent a "Field Safety Advisory Notice" dated 3/23/2017 to consignees. The notice described the product, problem and actions to be taken. The consignees were instructed to immediately examine the product and confirm correct product information; immediately discontinue use; move monitor suspension away from staff personnel and patients; and advising consignees to take photos of affected product and the P/N label. Photos of product labels are to be sent to jnewman@omegamedicalimaging.com at Omega. Product is to be corrected. If there are any questions or concerns, please feel free to contact us at 407-323-9400 Ext 203 or email: omegamedicalimaging.com or jnewman@omegamedicalimaging.com.

More recalls by Omega Medical Imaging, Inc.

View all recalls by this company →