2 over 2 Elevating Monitor Suspension - Omega Medical Imaging, LLC Elevating Monitor Suspension Systems.
This recall has been terminated (originally issued June 21, 2016).
- Company
- Omega Medical Imaging, Inc.
- Recall Initiated
- June 21, 2016
- Terminated
- October 4, 2019
- Recall Number
- Z-2734-2016
- Quantity
- 26 units
- Firm Location
- Sanford, FL
- Official Source
- View on FDA website ↗
Reason for Recall
Elevating Monitor Suspension separated from the lifting column bracket
Distribution
CO, MN, GA, NC, IN, OH, KY, IL, NY, AL, MO, LA, PA, and SD Denmark
Lot / Code Info
Model: Elevating Monitor Suuspension P/Ns: 1000-0085, 1000-0095 and (1000-0129 one unit produced does not use bracket 1040-0071 or 1040-0031).
Root Cause
Device Design
Action Taken
Consignees were notified by a Field Safety Advisory Notice - FSAN 16-0001 of this immediate Recall on 6/21/2016.