RecallDepth

2 over 2 Elevating Monitor Suspension - Omega Medical Imaging, LLC Elevating Monitor Suspension Systems.

This recall has been terminated (originally issued June 21, 2016).

Company
Omega Medical Imaging, Inc.
Recall Initiated
June 21, 2016
Terminated
October 4, 2019
Recall Number
Z-2734-2016
Quantity
26 units
Firm Location
Sanford, FL
Official Source
View on FDA website ↗

Reason for Recall

Elevating Monitor Suspension separated from the lifting column bracket

Distribution

CO, MN, GA, NC, IN, OH, KY, IL, NY, AL, MO, LA, PA, and SD Denmark

Lot / Code Info

Model: Elevating Monitor Suuspension P/Ns: 1000-0085, 1000-0095 and (1000-0129 one unit produced does not use bracket 1040-0071 or 1040-0031).

Root Cause

Device Design

Action Taken

Consignees were notified by a Field Safety Advisory Notice - FSAN 16-0001 of this immediate Recall on 6/21/2016.