Omega Medical Imaging Elevating Monitor Suspension, Part Number 1000-0085

This recall is currently active, issued October 24, 2023. It was issued by Omega Medical Imaging, Inc..

Company
Omega Medical Imaging, Inc.
Recall Initiated
February 28, 2023
Posted
October 24, 2023
Recall Number
Z-0156-2024
Quantity
52 units
Firm Location
Sanford, FL
Official Source
View on FDA website ↗

Reason for Recall

An actuator separated from the pivot mechanism on a lower monitor boom system column and became detached and was restrained by the installed safety cable at one location.

Distribution

Worldwide distribution - US Nationwide and the country of Pakistan.

Lot / Code Info

all units manufactured between 10-22-2012 and 07-09-2019

Root Cause

Device Design

Action Taken

Omega Medical Imaging issued a Field Safety Advisory Notice to its consignees on 02/28/2023 by email. The notice explained the issue and pledged to evaluate all potentially impacted devices and mitigate the issue to bring the boom systems up to the latest safety design and standard of the newer models approved for field use.