Medtronic Cardioblate Gemini-s, model 49260. Surgical Ablation Device. Sterilized using Ethylene Oxide. Two, unique jaw curvatures are provided: a standard curve (model 49260) and extra curve (Model 49261). The device is intended for intermittent operation. Manufacturer: Medtronic, Inc. Minneapolis, MN 55432-5604.

This recall has been terminated (originally issued February 2, 2008).

Company
Medtronic Cardiac Surgery Technologies
Recall Initiated
November 21, 2007
Posted
February 2, 2008
Terminated
December 17, 2011
Recall Number
Z-0727-2008
Quantity
28 units
Firm Location
Minneapolis, MN
Official Source
View on FDA website ↗

Reason for Recall

Mechanism failure: Medtronic has identified an issue with four lots of the Cardioblate Gemini-s Surgical Ablation Devices in which the polycarbonate distal coil retainer can fail during the course of a procedure, potentially resulting in an inability to apply the necessary jaw closure force.

Distribution

Nationwide Distribution including states of FL, IL, IN, LA, MI, PA, TN, TX, UT, and WI

Lot / Code Info

Lot Numbers: J243, J547, J548 and J656

Root Cause

Nonconforming Material/Component

Action Taken

Certified letters were sent on 11/21/07 to Consignees. The letter identifies the issue with the device and requests that customers return the device back to the firm. Customers will be requested to complete a Field Correction Action certificate acknowledging the receipt of the information. For additional information you can contact a Medtronic Sales Representative at 763-391-9167.