Medtronic U-CLIP Anastomotic Device Removal Tool, model REM1860FW. The U-Clip Anastomotic Device Removal Tool is designed to remove U-Clip Anastomotic Devices.

This recall has been terminated (originally issued June 27, 2007).

Company
Medtronic Cardiac Surgery Technologies
Recall Initiated
February 8, 2007
Posted
June 27, 2007
Terminated
February 24, 2008
Recall Number
Z-0968-2007
Quantity
187 (US 155, OUS 32)
Firm Location
Minneapolis, MN
Official Source
View on FDA website ↗

Reason for Recall

Six lots of U-Clip Removal Tools, catalog REM1860FW were incorrectly assembled. The nose cone attachment to the body of the removal tool does not contain sufficient adhesive which can cause the nose to separate from the body of the device.

Distribution

Nationwide and Canada.

Lot / Code Info

lot numbers: 0610707, 0614409, 0614410, 0614411, 0614412, 0622204

Root Cause

Other

Action Taken

Customers to be notified via a letter on 2/8/07. The letter will explain the situation and requests that the impacted product be returned to Medtronic for disposal. The customer will be requested to complete a "Recall Certificate" acknowledging the receipt of this information.