Medtronic SPYDER Anastomotic Device. Sterilized using irradiation. Do not Reuse. Nonpyrogenic. The SPYDER is designed to create an anastomosis in a rapid, automated fashion. The device creates an interrupted anastomosis without the need for aortic clamping. A specially designed aortic cutting device (ACD) (sold separately) is utilized to create the site for the anastomosis.

This recall has been terminated (originally issued April 13, 2006).

Company
Medtronic Cardiac Surgery Technologies
Recall Initiated
September 1, 2005
Posted
April 13, 2006
Terminated
April 1, 2007
Recall Number
Z-0763-06
Quantity
3383
Firm Location
Minneapolis, MN
Official Source
View on FDA website ↗

Reason for Recall

Customers stated that the U-Clip deployment collar was either difficult to rotate or 'locks up' during usage, which prevents the deployment of the U-Clips.

Distribution

Nationwide (AZ, CA, CT, FL, HI, IA, ID, IL, IN, KS, KY, LA, ,MD, MI, MN, MO, MS, NC, ND, NJ, NV, NY, OH, OK, PA, TN, TX, UT, WA, WI, WV). Japan, Belgium, Netherlands, Singapore, and Thailand.

Lot / Code Info

All units with Use Before Dates earlier than June, 2007.

Root Cause

Other

Action Taken

A letter and response certificate disseminated to all customers describing problem and recommending return of Spyder devices in inventory. September 1, 2005.