Medtronic SPYDER Anastomotic Device. Sterilized using irradiation. Do not Reuse. Nonpyrogenic. The SPYDER is designed to create an anastomosis in a rapid, automated fashion. The device creates an interrupted anastomosis without the need for aortic clamping. A specially designed aortic cutting device (ACD) (sold separately) is utilized to create the site for the anastomosis.
This recall has been terminated (originally issued April 13, 2006).
- Recall Initiated
- September 1, 2005
- Posted
- April 13, 2006
- Terminated
- April 1, 2007
- Recall Number
- Z-0763-06
- Quantity
- 3383
- Firm Location
- Minneapolis, MN
- Official Source
- View on FDA website ↗
Reason for Recall
Customers stated that the U-Clip deployment collar was either difficult to rotate or 'locks up' during usage, which prevents the deployment of the U-Clips.
Distribution
Nationwide (AZ, CA, CT, FL, HI, IA, ID, IL, IN, KS, KY, LA, ,MD, MI, MN, MO, MS, NC, ND, NJ, NV, NY, OH, OK, PA, TN, TX, UT, WA, WI, WV). Japan, Belgium, Netherlands, Singapore, and Thailand.
Lot / Code Info
All units with Use Before Dates earlier than June, 2007.
Root Cause
Other
Action Taken
A letter and response certificate disseminated to all customers describing problem and recommending return of Spyder devices in inventory. September 1, 2005.