VNUS Medical U-CLIP Removal Tool, UREM1860, is intended for the removal of U-Clip anastomotic devices.

This recall has been terminated (originally issued September 16, 2008).

Company
Medtronic Cardiac Surgery Technologies
Recall Initiated
February 25, 2008
Posted
September 16, 2008
Terminated
November 10, 2008
Recall Number
Z-1609-2008
Quantity
80
Firm Location
Minneapolis, MN
Official Source
View on FDA website ↗

Reason for Recall

Medtronic has identified one lot #0610315 of U-Clip Removal Tools, Catalog UREM1860FW, that was incorrectly assembled. The nose cone attachment, on the body of the removal tool, may not contain sufficient adhesive, which can allow the nose cone to separate from the body of the device.

Distribution

AZ, GA, PA, MI, KY, FL, OH, TX, NY, VA, CA, MO IL.

Lot / Code Info

Lot 0610315

Root Cause

Process control

Action Taken

A Medtronic "Urgent Medical Device Field Notification" letter was sent to one distributor and a separate consignee Medtronic letter "Urgent Medical Device Field Notification" was sent to end users on February 25, 2008. The letters explained the situation and requested that the impacted product be returned to Medtronic for disposal. The end user will be requested to complete a Recall certificate acknowledging the receipt of the information. If you have questions, contact your VNUS Sales Representative or Lauralie Millikan at 763-391-9124.