BD Biosciences MultiSET Flow Cytometry Software Versions 2.1, 2.1.1, and 2.2 with OS X operating systems, BD Biosciences, San Jose, CA 95131.

This recall has been terminated (originally issued March 4, 2008).

Company
Bd Biosciences
Recall Initiated
September 6, 2007
Posted
March 4, 2008
Terminated
April 15, 2008
Recall Number
Z-0570-2008
Quantity
1136 units
Firm Location
San Jose, CA
Official Source
View on FDA website ↗

Reason for Recall

Inaccurate results: Software error results in inaccurate display result statistics. If the user adjusts the lymph gate or attractors in the Lab Report view without using the Manual Gate function (as described in the system manual), the statistic results will not be updated to match the adjusted gate.

Distribution

Worldwide Distribution - USA, Belgium, Brazil, Peru, Singapore, Canada, Jamaica, Chile, Ecuador, Hong Kong-China, Thailand, Mexico, Venezuela, Barbados, Philippines, Puerto Rico, Guatemala, Japan, Uruguay, Argentina, China, Trinidad and Tobago, Honduras, Australia, New Zealand, and India.

Lot / Code Info

Software Versions 2.1, 2.1.1, and 2.2 with OS X operating systems.

Root Cause

Software design

Action Taken

Becton Dickinson sent an Important Product Information letter by mail on 09/14/2007 informing customers of the need to follow the instructions for adjusting the lymph gate or attractors. The next software revision will alleviate this limitation.