BD FACS Sample Prep Assistant II, Part Number: 337170; Product is manufactured and distributed by BD Biosciences, San Jose, CA 95131

This recall has been terminated (originally issued April 10, 2008).

Company
Bd Biosciences
Recall Initiated
October 2, 2007
Posted
April 10, 2008
Terminated
December 13, 2010
Recall Number
Z-0526-2008
Firm Location
San Jose, CA
Official Source
View on FDA website ↗

Reason for Recall

Mixing Performance: SPA II devices manufactured with the affected product motor control board may exhibit suboptimal mixing performance.

Distribution

Worldwide Distribution --- USA including states of: NJ, IL, TX, PA, NY, MN, and AZ, and countries of Australia, Belgium, Hong Kong, Trinidad and Tobago. Australia, Belgium, Hong Kong, Trinidad and Tobago.

Lot / Code Info

Serial Numbers: T0246 to T0290 (except for T0274,T0275,T0277 and T0279)

Root Cause

Software change control

Action Taken

On 10/1/07, all consignees were notified via Important Product Information letter and also an on-site visit by the firm's sales representative. The letter informed users of the affected product and provided instructions on the recall. For additional concerns or questions contact BD Customer Support Center at 1-877-232-8995.