RecallDepth

FACSDiva Software Versions 4.0.1 to 5.0.2, Flow Cytometry Acquisition & Analysis software. Catalog number 641609 Flexible data acquistion and analysis package specifically designed for digital-based flow cytometers.

This recall has been terminated (originally issued August 16, 2008).

Company
Bd Biosciences
Recall Initiated
April 5, 2007
Posted
August 16, 2008
Terminated
September 30, 2009
Recall Number
Z-1525-2008
Quantity
1074 units
Firm Location
San Jose, CA
Official Source
View on FDA website ↗

Reason for Recall

When data file containing one or no fluoresence parameters is exported, the software will automatically apply compensation to this file and all subsequently exported files.

Distribution

Worldwide Distribution --- USA including states of NY, NM, MI, GA, MD, MA, AL, CA, VA, NC, OH, TX, CO, NJ, PA, MI, CT, IA, MO, NV, KY, IL, TN, FL, MN, FL, UT, WA, OK, CT, WI, SD, AZ, and IN, and countries of Buenos Aires, BE, Brazil, Canada, Chile, China, Hong Kong, Japan, Singapore, Taiwan, Venezuela, Mexico, Trinidad and Tobago, and Columbia.

Lot / Code Info

Software versions 4.0.1 to 5.0.2.

Root Cause

Software design

Action Taken

The firm has issued a customer Important Product Information letter via mail on 04/20/2007 to inform them of the software issues. The letter states that to avoid this problem, make sure each tube or well has at least two fluorescence parameters before exporting FCS 3.0 data files. Manually exporting tubes or wells one at a time or restarting the software after the export of an FCS data file containing one or no fluorescence parameters will prevent the problem from occurring. Customers are requested to fill out and return the response form and to contact BD Customer Support Center if there are any additional questions. A software fix is being developed in version 6.0 and a free maintenance release with no product enhancments (version 5.0.3).