RecallDepth

XoranConnect, software version 4.1.0.6, Xoran Technologies, Ann Arbor, MI. A web-based software which provides online multi-planar image viewing, offsite data storage, administrative reporting tools, and teleradiology workflow. Server designed to store and serve stored imaging content on a RAID5 storage device.

This recall has been terminated (originally issued July 9, 2009).

Company
Xoran Technologies, Inc.
Recall Initiated
February 3, 2009
Posted
July 9, 2009
Terminated
November 19, 2009
Recall Number
Z-1577-2009
Quantity
14
Firm Location
Ann Arbor, MI
Official Source
View on FDA website ↗

Reason for Recall

An incorrect patient name label may appear on the screen displaying the selected patient's scan.

Distribution

Nationwide Distribution (Alabama, Arizona, Colorado, Georgia, Illinois, Louisiana, New York, North Carolina, Oklahoma, Rhode Island, Tennessee, Texas and Utah).

Lot / Code Info

Software version 4.1.0.6.

Root Cause

Software design

Action Taken

The firm released software version 4.1.0.7 by uploading it through a remote software upgrade on February 3, 2009. User accounts were notified by letter dated February 9, 2009. Direct questions to Xoran Technologies at 1-800-709-6726.

More recalls by Xoran Technologies, Inc.

View all recalls by this company →