Xoran Mini CT Scanner
This recall has been terminated (originally issued July 27, 2010).
- Company
- Xoran Technologies, Inc.
- Recall Initiated
- July 27, 2010
- Terminated
- September 11, 2014
- Recall Number
- Z-0036-2014
- Quantity
- 320
- Firm Location
- Ann Arbor, MI
- Official Source
- View on FDA website ↗
Reason for Recall
A description of the defect in the product or the manner in which the product fails to comply with an applicable Federal standard; The MiniCAT CT Scanners subject to this notification failed to meet Xoran's specifications for dose rate of 125 3 kVp. These 18 complaints received include dose rates both on the high and low end of Xoran's specifications. See ATTACHMENT 1 for details. (e) A n evalu
Distribution
Worldwide Distribution - US nationwide
Lot / Code Info
Minicat
Root Cause
Radiation Control for Health and Safety Act
Action Taken
All of the MiniCAT CT Scanners subject to these complaints were serviced by Xoran technicians and the out-of-specifications devices were brought into specification. This was accomplished by either readjusting the settings or by installing a new component.