RecallDepth

Xoran MiniCAT CT Scanner

This recall has been terminated (originally issued July 27, 2010).

Company
Xoran Technologies, Inc.
Recall Initiated
July 27, 2010
Terminated
September 11, 2014
Recall Number
Z-0037-2014
Quantity
320
Firm Location
Ann Arbor, MI
Official Source
View on FDA website ↗

Reason for Recall

A description of the defect in the product or the manner in which the product fails to comply wvith an applicable Federal standard; The MiniCAT CT Scanners subject to this notification failed to meet Xoran's specifications for dose rate of 125 3 kVp. These 18 complaints received include dose rates both on the high and low end of Xoran's specifications. See ATTACHMENT 1 for details. (e) A n evalu

Distribution

Worldwide Distribution - Nationwide

Lot / Code Info

Minicat

Root Cause

Radiation Control for Health and Safety Act

Action Taken

All of the MiniCAT CT Scanners subject to these complaints were serviced by Xoran technicians and the out-of-specifications devices were brought into specification. This was accomplished by either readjusting the settings or by installing a new component. Xoran's Research & Development group is evaluating the current specification to see if the field specifications are appropriate and to ensure that proper testing protocols are communicated and are in place for use by 3 rd party testing organizations. In addition, Xoran has initiated CAPA # 18 0 that is looking at the source block with regard to evaluating the stability of the dose rating.

More recalls by Xoran Technologies, Inc.

View all recalls by this company →