Agfa Heartlab Cardiovascular Results Management Product: Versions 2.03 and 2.04. Heartlab, Inc., Westerly, Rhode Island 02891

Company
Agfa Corp.
Recall Initiated
July 27, 2007
Posted
June 27, 2008
Terminated
October 10, 2008
Recall Number
Z-1242-2008
Quantity
6 units
Firm Location
Greenville, SC

Reason for Recall

Incorrect Results: Erroneous echocardiographic measurement values due to mathematical formula being mis-configured.

Distribution

Nationwide Distribution -- USA including states of GA, NH, NY and TX

Lot / Code Info

Results Management Model Number: L58RZ and L9MWQ, Version 2.03 and 2.04

Root Cause

Software design

Action Taken

On 7/27/2007 to 8/16/2007, AGFA Healthcare notified, via telephone, consignees of the situation. On 7/27/2007, a formal customer notification letter was sent to affected consignees, providing them with a respective listing of patients that were potentially affected. AGFA informed the consignees that the firm would correct any incorrect patient ECG data. AGFA also informed their consignees that they would provide a configuration update to correct the problem. A notification confirmation form to be return to the firm was also attached to the notification letter, which also provided the firm's customer service number: (877) 777-2432.