IMPAX CV Pediatric Echo Reporting, IMPAX CV Echo Measurement Import via Optical Character Recognition (OCR) Service IMPAX. For the creation of digital structured reports for cardiovascular diagnostic studies such as echocardiography.

Company
Agfa Corp.
Recall Initiated
April 30, 2014
Posted
May 27, 2014
Terminated
September 18, 2014
Recall Number
Z-1662-2014
Quantity
One
Firm Location
Greenville, SC

Reason for Recall

Failure to effectively install and test a hotfix to correct the known software defect within the IMPAX CV 7.8 SU3-OCR Service.

Distribution

Distribution to SC only

Lot / Code Info

Model L9M21000 - Software versions: IMPAX CV 7.8 SU3/SU4 - OCR Service

Root Cause

Software design

Action Taken

AGFA Healthcare sent an URGENT FIELD SAFETY NOTICE LETTER dated April 30, 2014, to the consignee describing the problem that had occurred, the actions being taken by Agfa and the actions Agfa has required the site to perform. As part of the correction Agfa requires the consignee to review the affected reports to confirm that the data is correct. If errors are found the consignee will need to correct the data. The consignee was asked to acknowledge, via FAX-Back to 864-421-1664 or email, that the information was received and understood by the consignee. For questions the consignee was instructed to contact their local Agfa HealthCare organization at 877-777-2432. For questions regarding this recall call 864-421-1754.