THD155428- 15.5F X 28cm Titan HD(TM) Catheter Kit Product Usage: Hemodialysis catheter insertion kit

Company
Medical Components, Inc Dba Medcomp
Recall Initiated
May 19, 2011
Posted
June 28, 2011
Terminated
October 24, 2016
Recall Number
Z-2693-2011
Quantity
90
Firm Location
Harleysville, PA

Reason for Recall

Product was packaged with the incorrect valved peelable introducer. The catheter sized is 15.5F and the introducer packaged in the kit is a 15F. The kit should contain a 16F introducer.

Distribution

Worldwide Distribution -- USA (nationwide) including states of AZ, CA, CO and RI and countries of Australia, Brazil, and Netherlands

Lot / Code Info

Lot # MAZX990; Catalog # THD155428, Exp Date 2013/03

Root Cause

Mixed-up of materials/components

Action Taken

Medcomp sent a "PRODUCT ALERT" letter dated May 19, 2011 to all affected customers. The letter described the product, problem, and actions to be taken. The letter is requesting the return of all affected product. Customers are instructed to contact their customer service representative for an RGA (returned good authorization) number. If you have any questions or concerns contact Medcomp at 215-256-4201 ext. 2225.