Power Injectable CT Port insertion kits - Product Usage: is an implantable access device designed to provide repeated access to the vascular system.

Company
Medical Components, Inc Dba Medcomp
Recall Initiated
January 13, 2021
Posted
March 8, 2021
Recall Number
Z-1184-2021
Quantity
96 pieces total (updated 3/11/2021 additional 72 pieces)
Firm Location
Harleysville, PA

Reason for Recall

The 5F Dignity CT Port kits were packaged with the incorrect size port. The kits include a 6.6F port instead of the correct 5F port. The inability to connect the catheter lumen to the port stem because of the incorrect size may not be noted until well into the procedure, after the catheter lumen is placed in the vessel. An alternate port will need to be obtained prolonging the procedure. A prolonged insertion procedure increases any risk associated with a surgical procedure.

Distribution

US Nationwide distribution in the states of CA, DC, GA, IL, IN, MA, MI, NJ, NM, NY, OH, PA, TX. Updated 3/11/2021

Lot / Code Info

Catalog Number: MRDP50AXS, Lot Number: MPDX180, UDI Number: 884908149753 (updated 3/11/2021 added lot number: MPCY660)

Root Cause

Employee error

Action Taken

On 1/13/2021 and 02/10/2021, Medical Components issued an Urgent Medical Device Removal notification to customers via letter notifying customers the 5F Dignity CT Port kits were packaged with the incorrect size port. The kits include a 6.6F port instead of the correct 5F port. (updated 3/11/2021)