Medcomp 15.SF X 24CM Titan HD CATHETER (CUFF 19CM FROM TIP) Ref: THD155024

Company
Medical Components, Inc Dba Medcomp
Recall Initiated
July 27, 2020
Terminated
April 30, 2021
Recall Number
Z-2940-2020
Quantity
132 eaches
Firm Location
Harleysville, PA

Reason for Recall

Kits were packaged with the incorrect size valve peelable introducer. The kits include a 15F instead of the correct 16F valve peelable introducer

Distribution

AZ, CA. FL. GA, NC, ND, NJ, NY, OH, PA, VA, WA

Lot / Code Info

Lot # MPCT670, UDI 884908129304 Exp. Date 12/20/2020

Root Cause

Under Investigation by firm

Action Taken

Medcomp issued Urgent Medical Device Recall letter issued first class mail to the customers on July 27, 2020. Letter states reason for recall, health risk and action to take: " Immediately examine your inventory and quarantine all product subject to recall. " Options for disposition of affected product: o Return the product to Medcomp - Contact your Medical Components, Inc. customer service representative for a Returned Goods Authorization {RGA) number. The phone number to call is 215-256-4201. o Have individually packaged and sterilized 16F peelable introducers (MR191603) sent to use with the kits. Call your customer service representative and provide the number of kits you need the replacements introducers for. Same number as above. " Complete and return the Medical Device Recall Product Disposition Form. The form can be returned by fax {215-256-9191) or email (recalls@medcompnet.com). Contact Information for questions: Monday through Friday, 8:00 AM to 4:00 PM, Eastern Standard Time. Phone: 215-256-4201; Email: recalls@medcompnet.com.

More recalls by Medical Components, Inc Dba Medcomp

View all recalls by this company →