RecallDepth

SoftMic Versions 4.0.4.7, 4.5.1.3. Version 4.0.4.7 Manufactured 02/03/2010. Version 4.5.1.3 Manufactured 03/18/2010. SCC Soft Computer 5400 Tech Data Drive Clearwater, FL 33760.

This recall has been terminated (originally issued October 19, 2010).

Company
Scc Soft Computer
Recall Initiated
July 19, 2010
Posted
October 19, 2010
Terminated
July 16, 2012
Recall Number
Z-0088-2011
Quantity
35
Firm Location
Clearwater, FL
Official Source
View on FDA website ↗

Reason for Recall

Softmic system may delay or omit reporting of clinically significant results including organisms and drug sensitivities. Incomplete culture results may appear complete. SCC Soft Computer is issuing a Correction for SoftMic Versions 4.0.4.7-4.5.1.3 software due to this potential hazard.

Distribution

Worldwide Distribution: USA, including the states of CA, CT, DE, FL, IA, KY, MA, MA, MI, MN, NJ, NY, OH, PA, TX, VA, and WI, and the country of Canada.

Lot / Code Info

Versions 4.0.4.7, 4.5.1.3.

Root Cause

Software design

Action Taken

SCC Soft Computer issued Risk-to-Health notifications to clients on July 19, 2010 via proprietary communication software. Corrective actions will be via programming correction via hot fix (software modification) or such so that Global variable is not used. Clients are asked to acknowledge receipt of this task and update the task with permission to load the required correction.

More recalls by Scc Soft Computer

View all recalls by this company →