SoftLab with SA HIS versions 4.0.7.0-4.0.7.1 SoftLab is a laboratory information system to be used in a medical research or clinical laboratory.

This recall has been terminated (originally issued February 2, 2015).

Company
Scc Soft Computer
Recall Initiated
May 20, 2014
Posted
February 2, 2015
Terminated
May 16, 2016
Recall Number
Z-1038-2015
Quantity
48
Firm Location
Clearwater, FL
Official Source
View on FDA website ↗

Reason for Recall

The interface fails to send abnormal flags for Reference Lab test results.

Distribution

Distributed US (nationwide) in the states of OH, GA, NJ, WI, FL, NH, IL, NY, IN, CA, AL, VA, MO, MI, KY, MA, PA, CO, ME, TN, WA, MN, SD, LA, NC, and NV and the country of Canada.

Lot / Code Info

4.0.7.0: released 12/13/13 - present 4.0.7.1: released 02/28/14 - present

Root Cause

Device Design

Action Taken

The firm sent out 2 CORRECTION COMMUNICATIONS informing consignees of functionality in release of 4.0.7 HIS Add-on interfaces which SCC has designated to be a potential patient Risk-to-Health issue. Further the corrective action indicated HIS releases 4.0.7.2 and higher will be corrected. Hot Fixes have been developed for HIS Add-on patches 4.0.7.0 and 4.0.7.1. Lastly the communication requests that the user "please acknowledge receipt of this task and grant permission to load the mandatory hot fix(es). Thank you."