RecallDepth

SoftPath Laboratory Information System. Versions 4.3.0.8, 4.3.0.9, 4.3.0.10, 4.3.0.11, 4.3.0.12, 4.3.0.14, 4.3.0.15, and 4.4.0.0

This recall has been terminated (originally issued February 9, 2015).

Company
Scc Soft Computer
Recall Initiated
November 26, 2014
Posted
February 9, 2015
Terminated
May 16, 2016
Recall Number
Z-1065-2015
Quantity
108 units
Firm Location
Clearwater, FL
Official Source
View on FDA website ↗

Reason for Recall

Modifications to diagnostic text may be: 1) Saved to the database but not appear on the report sent to the physician; or 2) Documented on the report, but not saved to the database.

Distribution

Distribution US Nationwide and CANADA.

Lot / Code Info

Version 4.3.0.8 Release date: 07/31/2009 - Present;  Version 4.3.0.9 Release date: 11/23/2009 - Present;  Version 4.3.0.10 Release date: 09/20/2010- Present;  Version 4.3.0.11 Release date: 11/11/2010- Present;  Version 4.3.0.12 Release date: 02/07/2011- Present; Version 4.3.0.14 Release date: 11/14/2012- Present;  Version 4.3.0.15 Release date: 09/17/2013- Present;  Version 4.4.0.1 Release date: 12/13/2013 - Present;

Root Cause

Software design

Action Taken

On 11/26/2014 SCC Soft Computer sent a Correction Communication to their consignees: Corrective Action: Versions 4.3.0.8 through 4.4.0.2 will have a mandatory hotfix created. Due to the amount of versions affected, the range of availability dates will be 11/24/2014 through 12/30/2014. Utility: A utility will be made available to be run on your Live system that will identify cases in which there is a disparity between the diagnosis text in the document and the database. The utility output will be provided for client review and validation.