RecallDepth

MicroVention Terumo Microplex 10, Platinum Coil System, Endovascular Embolization Coil, Compass 10, Sterile, Rx, REF numbers 100203CMSR-V, 100408CMSR-V, 100412CMSR-V, 100510CMSR-V, 100516CMSR-V, 100612CMSR-V, 100618CMSR-V, and 100721CMSR-V. Product Usage: The MCS is also intended for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolizations in the peripheral vasculature. The firm name on the label is MicroVention, Inc., Tustin, CA.

Company
Microvention Inc.
Recall Initiated
November 22, 2019
Terminated
January 26, 2022
Recall Number
Z-1404-2020
Quantity
85 units
Firm Location
Aliso Viejo, CA

Reason for Recall

The devices may be missing the implant coil.

Distribution

All distribution was OUS. The countries receiving affected product are: Albania, Austria, Belarus, Belgium, Brazil, Bulgaria, China, Cyprus, Czech Republic, Denmark, Egypt, Finalnd, France, Georgia, Germany, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Korea, Latvia, Lebanon, Malta, Morocco, Netherlands, Norway, Poland, Portugal, Republic of Serbia, Romania, Russian Federation, Slovak Republic, Slovenia, South Africa, Spain, Sweden, Switzerland, Syria, Turkey, Ukraine, United Arab Emirates, and United Kingdom.

Lot / Code Info

Catalog/lot numbers:  100203CMSR-V - 1903085WP;  100408CMSR-V - 1903085WP and 1904015PP;  100412CMSR-V - 1904225RP, 1905205QP, and 1904035PP;  100510CMSR-V - 1905205QP and 1904085WP;  100516CMSR-V - 1903135WP and 1905155WP;  100612CMSR-V - 1903225WP;  100618CMSR-V - 1903135QP and 1904085WP; and   100721CMSR-V - 1903085RP.

Root Cause

Process design

Action Taken

The recalling firm issued letters beginning 11/22/2019 dated 11/26/2019 via email using different flags according to the region the foreign consignee was located, i.e. 'URGENT: FIELD SAFETY CORRECTION ACTION" (for EMEA), "URGENT: MEDICAL DEVICE RECALL" (for Japan), or "FIELD SAFETY ADVISORY NOTICE" (for Rest of World).

More recalls by Microvention Inc.

View all recalls by this company →