Flow Re-Direction Endoluminal Devices: FRED 27, REF: FRED5019, FRED5029, FRED4528, FRED4026, FRED5526, FRED4508, FRED5514, FRED3536, FRED3511, FRED5014, FRED3516, MV-F451827; FRED X 27, REF: XFRED4017-MVE, XFRED4528-MVE, MV-F501427X, XFRED4518-MVE, XFRED5526-MVE, XFRED4539-MVE, FREDX5522-PMA, FREDX3522-PMA, FREDX4525-PMA, FREDX5015-PMA, FREDX5532-PMA, FREDX4013-PMA, FREDX3513-PMA, FREDX4515-PMA, FREDX3517-PMA, MV-F352427X

Company
Microvention Inc.
Recall Initiated
September 5, 2025
Posted
October 10, 2025
Recall Number
Z-0108-2026
Quantity
467
Firm Location
Aliso Viejo, CA

Reason for Recall

Flow diverter may have a manufacturing issue related to the tantalum length and/or tantalum attachment pattern, which may lead the device to be incompletely open or not properly apposed to the vessel wall, which may necessitate device recapture and withdrawal of the delivery system and microcatheter together. Insufficient apposition could lead to long-term health consequences.

Distribution

Worldwide - US Nationwide distribution in the states of SD, CA, SC, TX, FL, CO, NE, NY and the countries of DE, ES, PL, GB, IT, IN, AU, KZ, JP, AT, SI, RS, GR, TR, AR, CH, SG, KR, SK, MX, DK, IR, SA, BY, PA, VN, ZA, BR, TW, CN, BE, FR, VE, MY.

Lot / Code Info

REF/UDI-DI(Lot): FRED5019/00811425020388(0000235839), 00842429117163(0000544498); FRED5029/00811425020395(0000256343), 00842429117170(0000507209); FRED4528/00811425020425(0000265666); FRED4026/00811425021033(0000333125, 0000576973); FRED5526/00842429117194(0000361117); FRED4508/00842429117095(0000372088); FRED5514/00842429117187(0000429786, 0000478385); FRED3536/00842429117033(0000460268, 0000445555); FRED3511/00842429117002(0000452923); FRED5014/00842429117156(0000563591, 0000464401); FRED3516/00842429117019(0000513146, 0000531733, 0000536217); MV-F451827/04987892121906(0000565094); XFRED4017-MVE/00842429107508(0000478838); XFRED4528-MVE/00842429107560(0000518323, 0000663751); MV-F501427X/04987892151231(0000579031); XFRED4518-MVE/00842429107553(0000663074, 0000663750); XFRED5526-MVE/00842429107638(0000704203); XFRED4539-MVE/00842429107577(00007311620); FREDX5522-PMA/00842429114483(0000265641, 0000469609, 0000731155); FREDX3522-PMA/00842429114315(0000469598); FREDX4525-PMA/00842429114414(0000469603); FREDX5015-PMA/00842429114445(0000469606); FREDX5532-PMA/00842429114490(0000469610); FREDX4013-PMA/00842429114346(0000469613, 0000663758, 0000668189); FREDX3513-PMA/00842429114292(0000478834); FREDX4515-PMA/00842429114391(0000478840); FREDX3517-PMA/00842429114308(0000731304); MV-F352427X/04987892151101(0000760963)

Root Cause

Device Design

Action Taken

On 9/5/2025, recall notices were mailed to customers who were asked to do the following: 1) Stop using and quarantine affected devices. 2) Provide the notice to medical facilities or users to whom you have distributed recalled devices. 3) Return effected devices. 4) Complete and return the customer acknowledgement form via email to Terumo8943@sedgwick.com Direct questions to the firm at terrence.callahan@microvention.com On 10/29/2025, additional recall notices were distributed identifying additional distributed devices.