RecallDepth

Microvention, Headway" 17 Advanced Straight Microcatheter with Hydrophilic Coating, REF: MC172150SX-CN, Sterile R, Rx Only, Lot# 0000629846

Company
Microvention Inc.
Recall Initiated
December 2, 2024
Posted
January 17, 2025
Recall Number
Z-0990-2025
Quantity
107 units
Firm Location
Aliso Viejo, CA

Reason for Recall

Sterility assurance; microcatheter inner packaging may contain incomplete seal.

Distribution

US: None OUS: China

Lot / Code Info

Lot # 0000629846/UDI: (01)00816777027634

Root Cause

Packaging process control

Action Taken

On December 11, 2024, MicroVention Eurpose SARL dba Terumo Neuro intiated a "Urgent Medical Device Recall" via mail. Terumo Neuro asked consignees to take the following actions: 1. Review Inventory 2. Inform and forward recall notice 3. Return Product 4. Contact Quality Department 5. Replacement or Credit of returned devices 6. Adverse reaction or quality problems experienced with the use of this product may be reported to your local MicroVention Terumo Sales Representative or the FDA's MedWatch Adverse Event Reporting program either online, by regular mail or by fax.

More recalls by Microvention Inc.

View all recalls by this company →