Ratcheting Cannulated Driver Handle, Catalog #: 210-00-004, Rx Only, Non Sterile

This recall has been terminated (originally issued March 27, 2018).

Company
Trilliant Surgical Ltd.
Recall Initiated
March 27, 2018
Terminated
June 4, 2020
Recall Number
Z-2339-2018
Quantity
90
Firm Location
Houston, TX
Official Source
View on FDA website ↗

Reason for Recall

Firm is voluntarily recalling cannulated driver handles (210-00-003 and 210-00-004) that may demonstrate a degradation of function with the adapter collar sticking, and/or not retaining the mating adapter male shaft.

Distribution

U.S. States: NE, MA, GA, TX, OR, AZ, FL, OH, WA, NJ, UT, VA, CA, MN, PA, DE, OK, KS, WI, MT, IA, MO

Lot / Code Info

Lot: 603600

Root Cause

Other

Action Taken

On 03/27/18, the firm sent Medical Device Recall notices to customers informing them to: 1) Check inventory stock for identified devices; 2) Locate and remove and/or discontinue use of identified devices; 3) Return all related product(s) to the firm in accordance with an included return response / acknowledgement form. Customers were instructed to contact the firm at 1-800-495-2919 to organize product returns.