7 Hole VL Gridlock Fibula Plate, Part # 300-60-001 NON-STERILE SINGLE USE ONLY; 9 Hole VL Gridlock Fibula Plate, Part # 300-60-002 NON-STERILE SINGLE USE ONLY; 11 Hole VL Gridlock Fibula Plate, Part # 300-60-003 NON-STERILE SINGLE USE ONLY. For use in trauma and reconstructive procedures in small bones.

This recall has been terminated (originally issued August 31, 2015).

Company
Trilliant Surgical Ltd.
Recall Initiated
August 6, 2015
Posted
August 31, 2015
Terminated
January 26, 2017
Recall Number
Z-2490-2015
Quantity
28 units
Firm Location
Houston, TX
Official Source
View on FDA website ↗

Reason for Recall

The affected parts subject to the recall are out of specification, resulting in in the loss of the optimal locking screw functionality of the plates. The thread depth of the screw holes is insufficient, potentially limiting the ability to engage and lock with associated mating locking screws.

Distribution

US Distribution to the states of : FL, OH, NJ, MN, WI, NM and CA.

Lot / Code Info

Lots: TSL002595, TSL002596, TSL002597

Root Cause

Nonconforming Material/Component

Action Taken

Trilliant Surgical sent an Urgent Medical Device Recall letter dated August 13, 2015, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The letter instructed and confirmed consignees to discontinue use of the affected lots, quarantine, and return product (if not yet done so) in accordance with an included acknowledgement sheet. For further questions, please call 1(800) 495-2919 , 8:00am - 5:00pm CST