Tiger Screw Sterilization Tray Base Assembly Model 210-00-001 Revision N, O The product is a component of a system indicated for the fixation of fractures, non-unions, arthrodeses, and osteotomies of bones appropriate for the size of the device.

This recall has been terminated (originally issued January 6, 2017).

Company
Trilliant Surgical Ltd.
Recall Initiated
December 8, 2016
Posted
January 6, 2017
Terminated
May 17, 2017
Recall Number
Z-0960-2017
Quantity
50 units
Firm Location
Houston, TX
Official Source
View on FDA website ↗

Reason for Recall

Four lots of Tiger Screw Sterilization Tray Base Assemblies were released to the field with a printing error where k-wire description markings indicate diameters of 0.45 and 0.35 instead of 0.045" and 0.035", respectively.

Distribution

Nationwide Distribution

Lot / Code Info

Lot: TSL003191, TSL003637, TSL004101, TSL004395

Root Cause

Under Investigation by firm

Action Taken

Trilliant Surgical Ltd. sent an Urgent Medical Device Recall letter dated December 8, 2016, to all affected consignees. The letter instructed the consignees to 1) identify any/all affected product; 2) document any/all identified product via a provided response form and return to Trilliant Surgical; 3) exchange product upon receipt of replacements; and 4) return any/all affected product as originally identified and reported on the return response form. Customers with questions were instructed to call 800-495-2919.