Torque Limiting Handle: PRECIMED, T-Handle Torque, Non Sterile. Handle: Customer Number/Model: MCL120014 T14000, TLT911-50-S20 TLT911-50-S20, TLT911-15-S01 TLT911-15-S01, TLTS0007-20-S19 TLTS0007-20-S19, TLT911-35-S01 TLT911-35-S01.

This recall has been terminated (originally issued April 26, 2016).

Company
Greatbatch Medical
Recall Initiated
April 26, 2016
Terminated
June 7, 2018
Recall Number
Z-1882-2016
Quantity
2
Firm Location
Plymouth, MN
Official Source
View on FDA website ↗

Reason for Recall

Greatbatch Medical has initiated a global Field Recall for Torque Limiting Devices. Greatbatch identified through laboratory testing that the required sterility assurance level (SAL) of 10-6 is not achieved on torque limiting devices when subjected to steam sterilization cycles identified in the Instruction For Use (IFU) provided with the device.

Distribution

US: IN, MI, TX. OUS: GERMANY, FRANCE, ITALY, SWITZERLAND.

Lot / Code Info

6307760001

Root Cause

Device Design

Action Taken

Consignees were sent on 4/26/2016 a Greatbatch "Urgent Medical Device Recall" letter dated April 26, 2016. The letter described the product involved in the recall, the reason, and the risk to health. The letter advised consignees to identity and quarantine the product and to complete and return the Filed Recall Response Form. For questions contact your local Greatbatch Medical representative, at 1-763-951-8235, or e-mail FieldActionCenter@Greatbatch.com.