ViaPeel PTFE Peelable Introducer 5FR, Part Number 10890-006

This recall has been terminated (originally issued November 11, 2016).

Company
Greatbatch Medical
Recall Initiated
November 11, 2016
Terminated
November 3, 2017
Recall Number
Z-2781-2017
Quantity
5,000 devices
Firm Location
Plymouth, MN
Official Source
View on FDA website ↗

Reason for Recall

Greatbatch received complaints with the ViaPeel Introducer Sheath 5FR, of handles detaching during use.

Distribution

NY

Lot / Code Info

Lot Number: W3101005

Root Cause

Process control

Action Taken

Consignee was sent via email on 11/11/2016 a Greatbatch "Urgent Medical Device Recall" letter dated November 2016. The letter described the product involved in the recall, Reason for Field Corrective Action, Risk to Health, Replacement Devices & Instructions to Customer. Advised consignees to quarantine the product, contact end users & complete and return the Field Recall Response Form to Stericycle at Greatbatch7893@stericycle.com or fax at 888-473-8015. For questions contact your local Greatbatch Medical representative, contact us at 1-763-951-8235, or e-mail us at FieldActionCenter@Greatbatch.com