Greatbatch Stiffer Coaxial Micro-Introducer. Model #/ Name: 10904-001 Stiffer Coaxial Micro-Introducer 4FR; 10904-002 Stiffer Coaxial Micro-Introducer 5FR. The radiopaque Stiffer Coaxial Micro-Introducer (Figure 1) consists of a dilator and sheath, available in 4FR and 5FR. The inner dilator includes a stainless steel stiffening hypotube. The Coaxial Micro-Introducer is indicated for percutaneous introduction of up to a 0.038 in. guidewire or catheter into the vascular system through an initial puncture of a 21G introducer. The device is sold to customers in a bulk non-sterile configuration.

This recall has been terminated (originally issued November 18, 2016).

Company
Greatbatch Medical
Recall Initiated
September 8, 2016
Posted
November 18, 2016
Terminated
September 8, 2017
Recall Number
Z-0623-2017
Quantity
23,075
Firm Location
Plymouth, MN
Official Source
View on FDA website ↗

Reason for Recall

Greatbatch Medical has identified an issue with the Stiffer Coaxial Micro-Introducer 4FR and 5FR after a field complaint reported that the rotating luer detached from the dilator hub during use.

Distribution

US: NY, TX.

Lot / Code Info

GBP3176925, GBP3176927, VN3470927, VN3470929, VN3486000, VN3485998, VN3495047, GBP3421664, VN3433259, VN3451654, VN3487654, VN3491069, VN3470925, VN3451646.

Root Cause

Process design

Action Taken

Greatbatch sent a "Urgent Medical Device Recall" letter dated September 7, 2016 to all affected customers .The letter described the product, "Reason for Field Correction Action", "Risk to Health" , "Replacement Devices", & "Instructions to Customers" . Advised customers to quarantine the product and contact end users and share the recall information with them. Requested consignees to complete & return the Field Recall Response Form. For questions contact your local Greatbatch Medical representative, contact us at 1-763-951-8235, or e-mail us at FieldActionCenter@Greatbatch.com.