RayStation 3.5, RayStation 4.0, RayStation 4.5, RayStation 4.7, RayStation 5, RayStation 6, RayPlan 1 (RayStation 4.9) and RayPlan 2 (RayStation 6) . Intended Use: software system designed for treatment planning and analysis of radiation therapy

Company
Raysearch Laboratories Ab Sveavagen 9 Stockholm Sweden
Recall Initiated
November 20, 2017
Terminated
August 8, 2023
Recall Number
Z-1930-2019
Quantity
744

Reason for Recall

When calculating electron Monte Carlo dose with a very large number of histories, the dose calculation may be wrong

Distribution

US Nationwide Distribution

Lot / Code Info

Software versions:  3.5.0.16 , 3.5.1.6 , 4.0.0.14 , 4.0.1.4 , 4.0.2.9 , 4.0.3.4 , 4.5.0.19 , 4.5.1.14 , 4.5.2.7 , 4.7.0.15 , 4.7.1.10 , 4.7.2.5 , 4.7.3.13 , 4.7.4.4 , 4.7.5.4 , 4.9.0.42 , 5.0.0.37 , 5.0.1.11 , 5.0.2.35 , 5.0.3.17 , 6.0.0.24 , 6.1.0.26 , 6.1.1.2 and 6.2.0.7

Root Cause

Software design

Action Taken

RaySearch issued Field Safety Notice, Medical Device Correction #24138 via email on November 20th, 2017. The letter identifies the product, issue, health risk and action to take: Do not use a number of histories above 6 million for electron dose calculation. Inspect all commissioned electron beam models for any applicator calculated with greater than 6 million. Contact customer support if you have any further questions. Confirm you have read and understood this notice. This issue will be resolved in the next version of RayStation, scheduled for market release in December 2017. Educate planning staff and all users about this workaround.