RayStation 11B

Company
Raysearch Laboratories Ab Sveavagen 9 Stockholm Sweden
Recall Initiated
June 10, 2022
Terminated
June 11, 2024
Recall Number
Z-1299-2022
Quantity
26

Reason for Recall

This notice concerns two issues found related to the display of Linear Energy Transfer (LET) in RayStation 11B including some service packs. First, when using a dose threshold for an evaluation dose, LET display might be misleading. Second, a displayed beam-specific LET distribution can sometimes be out of sync with the selected beam.

Distribution

US Nationwide distribution in the states of NJ and TX.

Lot / Code Info

UDI-DI 0735000201042620211208 Software version numbers 12.0.0.932, 12.1.0.1221, 12.0.3.68

Root Cause

Under Investigation by firm

Action Taken

The firm notified customers of the recall via email on June 10, 2022. The notice was titled "Field Safety Notice, Medical Device Correction #99834." Customers were given precautions and instructed on a workaround. The issue will be resolved in the next version of RayStation, scheduled for market release in June 2022, though this may be subject to market clearance.