Concave Reamer Head, 16 mm; Concave Reamer Head, 18 mm; Concave Reamer Head, 20 mm; Concave Reamer Head 22 mm; Convex Reamer Head, 16 mm; Convex Reamer Head, 18 mm; Convex Reamer Head 20 mm; Convex Reamer Head 22 mm, used with Gridlock Plating system, for use in trauma and reconstructive procedures in small bones.

This recall has been terminated (originally issued August 24, 2015).

Company
Trilliant Surgical Ltd.
Recall Initiated
July 20, 2015
Posted
August 24, 2015
Terminated
January 26, 2017
Recall Number
Z-2456-2015
Quantity
638 units
Firm Location
Houston, TX
Official Source
View on FDA website ↗

Reason for Recall

The cannulation of the Concave and Convex Reamer Heads may be too small to allow the Guide Pin to freely pass through.

Distribution

Nationwide throughout the US

Lot / Code Info

TSL000145, TSL000162, TSL000699, TSL000146, TSL000163, TSL000700, TSL00147, TSL000164, TSL000701, TSL000702, TSL000148, TSL000165, TSL000706, TSL000149, TSL000166, TSL000688, TSL000150, TSL000167, TSL000707, TSL000708

Root Cause

Component design/selection

Action Taken

The recalling firm sent out written notification of the recall on 7/3/14. The letter instructed the consignees to exchange affected instruments.