IMPAX CV Reporting Cardiac Catheterization module and IMPAX CV Outbound Reporting. Facilitates digital reports for catheterization, echocardiography, and congenital heart disease.

Company
Agfa Corp.
Recall Initiated
March 15, 2013
Posted
April 8, 2013
Terminated
January 9, 2014
Recall Number
Z-1070-2013
Quantity
26
Firm Location
Greenville, SC

Reason for Recall

Baseline pulmonary capillary wedge (PCW) pressure values from IMPAX CV Reporting Cardiac Catheterization module were not correctly transferring into the hospital's electronic health record (EHR) system.

Distribution

Nationwide Distribution including the states of AK, GA, IL, KY, MA, NY, NC, RI, SC, TX, and VA.

Lot / Code Info

Model # L9M21000. Software versions: IMPAX CV 7.8, 7.8 SU1, 7.8 SU2 and 7.8 SU3, IMPAX CV 7.8 SU2.5 with IMPAX CV Outbound Report Stylesheet 7.8_HL

Root Cause

Software design

Action Taken

AGFA sent an "URGENT FIELD SAFETY NOTICE" letter dated March 15, 2013 to the consignees who use the IMPAX CV Reporting tool described for this issue. The letter identified the product, problem, and actions to be taken by the customers. Contact the firm at 877-777-2432, referencing HQ_10020006 for questions regarding this notice.